The 3/8-inch Venous Line: Conserving Blood, but Not Preserving It?

Mya Rai
DAY 1
DAY 2
DAY 3

An in vitro evaluation of venous tubing calibre and the addition of vacuum assisted venous drainage as determinants of haemolysis during simulated cardiopulmonary bypass.
‍

Abstract
‍

Background. Haemodilution from the cardiopulmonary bypass (CPB) prime is a modifiab factor associated with patient outcomes, and circuit tubing contributes substantially to prime volume. Narrowing the venous line from 1/2-inch to 3/8-inch lowers venous line volume by approximately 44% but raises flow velocity, wall shear stress and hydraulic resistance, potentially requiring vacuum-assisted venous drainage (VAVD). Increased mechanical stress and VAVD have been implicated in red blood cell (RBC) injury, so a strategy intended to limit haemodilution may damage the cells it is designed to conserve.
‍

Aim. To determine whether reducing venous calibre from 1/2-inch to 3/8-inch increases haemolysis during simulated CPB, whether VAVD magnitude does so independently, and whether the two interact.
‍

Methods. A closed-loop benchtop CPB circuit was run with porcine erythrocytes at a haematocrit of 30%. Venous calibre (3/8-inch versus 1/2-inch) was crossed with four VAVD
levels (0, 20, 40 and 60 mmHg), giving eight two-hour runs at 5 L/min. Blood was sampled at five time points and plasma free haemoglobin (PfHb) measured, giving 40 observations, analysed by linear mixed-effects regression with a random intercept for run.
‍

Results. No difference in ΔPfHb was detected between the 3/8-inch and 1/2-inch lines (p =0.715), across VAVD levels (p = 0.708), or for the calibre × VAVD interaction (p = 0.398). Circulation time was the only predictor for which an association was detected (β = 0.036 per minute on the ln(ΔPfHb + 1) scale, 95% CI 0.029 to 0.043; p < 0.001).
‍

Conclusion. Reducing venous calibre and applying moderate VAVD produced no statistically detectable increase in overt haemolysis under the conditions tested. However, the unreplicated experimental design limits the reliability of these findings, and sublytic RBC injury cannot be excluded. The safety of the narrower line therefore remains unestablished in the model and requires further investigation.

SPEAKER PROFILE

Mya Rai is a Clinical Perfusionist at Basildon Hospital with a BSc in Physiology from University College London and an MSc in Perfusion Science from the University of Bristol. She has a particular interest in clinical research and evidence-based practice.

‍

VIEW ABSTRACTS BY OTHER SPEAKERS

Ryan Sapiandante
GUEST
Ben Fryer
GUEST
Karl Riley
GUEST
Kirsty Dempster
GUEST
Saeed Taiyari
GUEST
Matthew Donald
GUEST
Joseph Hughes
GUEST
Leigh-Anne Brooks
GUEST
Mike Adamek
GUEST
Peter O'Sullivan
GUEST
Michael Van Driel
SPONSOR
London Core Review Cardiothoracic Surgery Course - Touch icon
London Core Review Cardiothoracic Surgery Course - Help Guide
SPONSORS